Clinical trialNo phase (not a drug trial)Completed
Malignant Obstruction ZILVER Against Routine Therapy (MOZART I)
Multicenter Prospective Randomized Controlled Trial of the Nitinol ZILVER Expandable Endoprosthesis in the Palliation of Extrahepatic Malignant Biliary Obstruction
- Outcome
- Primary endpoint: 3 of 8 posted analyses statistically significant
- Phase
- No phase (not a drug trial)
- Status
- Completed
- Study type
- Interventional
- Start
- September 2003
- Ended
- December 2006
- Enrollment
- 241 participants
- Sites
- 1
- Primary outcomes
- Patency; Closure or Blockage of the Stent (Occlusion); Number of Days to Occlusion; Number of Deaths; Time to Death
- Sponsor type
- Industry
- Results posted
- 12 January 2010
- Last update
- 3 October 2011
These fields come from the trial’s registry entry. A summary of its population, design, endpoints and results, each tied to a quote, has not been written for this trial yet.
Results
- Primary endpoint
- Primary endpoint: 3 of 8 posted analyses statistically significantAs posted on ClinicalTrials.gov ↗, registry record of 3 October 2011
- Primary outcome
- Patency6 mm Zilver: 110 (7 to 595), n = 6410 mm Zilver: 125.5 (5 to 671), n = 8810 mm Wallstent: 121 (0 to 961), n = 896 mm Zilver vs 10 mm Zilver vs 10 mm Wallstent: p = .057; Log Rank (not statistically significant against 0.05 (no threshold posted); outcome 1, analysis 1)Values: median Days, Full RangeTime frame: up to 32 months
- Primary outcome
- Closure or Blockage of the Stent (Occlusion)6 mm Zilver: 25, n = 6410 mm Zilver: 21, n = 8810 mm Wallstent: 19, n = 896 mm Zilver vs 10 mm Zilver vs 10 mm Wallstent: p = .04; Chi-squared (statistically significant against 0.05 (no threshold posted); outcome 2, analysis 1)6 mm Zilver vs 10 mm Zilver vs 10 mm Wallstent: p = .007; Log Rank (statistically significant against 0.05 (no threshold posted); outcome 2, analysis 2)10 mm Zilver vs 10 mm Wallstent: p = .69; Chi-squared (not statistically significant against 0.05 (no threshold posted); outcome 2, analysis 3)6 mm Zilver vs 10 mm Zilver vs 10 mm Wallstent: p = .02; Chi-squared (statistically significant against 0.05 (no threshold posted); outcome 2, analysis 4)Values: Stent OcclusionsTime frame: up to 32 months
- Primary outcome
- Number of Days to Occlusion6 mm Zilver: 115 (7 to 595), n = 2510 mm Zilver: 111 (6 to 506), n = 2110 mm Wallstent: 103 (7 to 386), n = 196 mm Zilver vs 10 mm Zilver vs 10 mm Wallstent: p = .16; ANOVA (not statistically significant against 0.05 (no threshold posted); outcome 3, analysis 1)Values: median Days, Full RangeTime frame: up to 32 months
- Primary outcome
- Number of Deaths6 mm Zilver: 4, n = 6410 mm Zilver: 6, n = 8810 mm Wallstent: 2, n = 89Values: ParticipantsTime frame: up to 32 monthsNo analysis posted for this outcome (outcome 4)
- Primary outcome
- Time to Death6 mm Zilver: 132.6 (74 to 328), n = 6410 mm Zilver: 186 (92 to 304), n = 8810 mm Wallstent: 170 (99 to 273), n = 896 mm Zilver vs 10 mm Zilver vs 10 mm Wallstent: p = .32; Log Rank (not statistically significant against 0.05 (no threshold posted); outcome 5, analysis 1)6 mm Zilver vs 10 mm Zilver vs 10 mm Wallstent: p = .69; Kruskal-Wallis (not statistically significant against 0.05 (no threshold posted); outcome 5, analysis 2)Values: median Days, Inter-Quartile RangeTime frame: up to 32 months
- Participants
- 6 mm Zilver: 64 started, 60 completed10 mm Zilver: 88 started, 83 completed10 mm Wallstent: 89 started, 82 completed
- Serious adverse events
- 6 mm Zilver: 0 of 010 mm Zilver: 0 of 010 mm Wallstent: 0 of 0Participants with at least one serious adverse event, of those at risk, as posted
- Reported in
- Comparative performance of uncoated, self-expanding metal biliary stents of different designs in 2 diameters: final results of an internati… DOI 10.1016/j.gie.2008.11.018↗ Linked as this trial’s results paper. The numbers above are the registry’s, not the paper’s.
Every value is copied from the trial’s results record at ClinicalTrials.gov. “Statistically significant” describes the posted test only: it does not say which group did better, and nothing here is read from a paper.
Ask this document
Answers come only from this document’s own text, as highlighted quotes. If it doesn’t say, you are told so instead of given a guess.
Try
Source text
Brief summary public domain (ClinicalTrials.gov)
This study compares the 6 mm nitinol Zilver biliary endoprostheses and the 10 mm nitinol Zilver biliary endoprostheses to the 10 mm Wallstent in appropriate patients in need of palliative treatment of malignant obstructive jaundice.
Detailed description public domain (ClinicalTrials.gov)
This research is being done to determine if the new, FDA-cleared Zilver metal biliary stent is better than the conventional stents. The new Zilver stent may permit more accurate placement, may avoid some potential complications and last longer than the conventional stents. The study will also examine to find the best diameter of stent by comparing standard 10 mm stents to 6 mm Zilver stents.
Posted results (ClinicalTrials.gov registry) · primary outcome public domain (ClinicalTrials.gov)
Patency [up to 32 months]. Median, Days (Full Range). 6 mm Zilver (n=64): 110 (7 to 595); 10 mm Zilver (n=88): 125.5 (5 to 671); 10 mm Wallstent (n=89): 121 (0 to 961). Analysis, 6 mm Zilver vs 10 mm Zilver vs 10 mm Wallstent: p=.057; Log Rank; superiority or other.
Posted results (ClinicalTrials.gov registry) · primary outcome public domain (ClinicalTrials.gov)
Closure or Blockage of the Stent (Occlusion) [up to 32 months]. Number, Stent Occlusions. 6 mm Zilver (n=64): 25; 10 mm Zilver (n=88): 21; 10 mm Wallstent (n=89): 19. Analysis, 6 mm Zilver vs 10 mm Zilver vs 10 mm Wallstent: p=.04; Chi-squared; superiority or other. Analysis, 6 mm Zilver vs 10 mm Zilver vs 10 mm Wallstent: p=.007; Log Rank; superiority or other. Analysis, 10 mm Zilver vs 10 mm Wallstent: p=.69; Chi-squared; superiority or other. Analysis, 6 mm Zilver vs 10 mm Zilver vs 10 mm Wallstent: p=.02; Chi-squared; superiority or other.
Posted results (ClinicalTrials.gov registry) · primary outcome public domain (ClinicalTrials.gov)
Number of Days to Occlusion [up to 32 months]. Median, Days (Full Range). 6 mm Zilver (n=25): 115 (7 to 595); 10 mm Zilver (n=21): 111 (6 to 506); 10 mm Wallstent (n=19): 103 (7 to 386). Analysis, 6 mm Zilver vs 10 mm Zilver vs 10 mm Wallstent: p=.16; ANOVA; superiority or other.
Posted results (ClinicalTrials.gov registry) · primary outcome public domain (ClinicalTrials.gov)
Number of Deaths [up to 32 months]. Number, Participants. 6 mm Zilver (n=64): 4; 10 mm Zilver (n=88): 6; 10 mm Wallstent (n=89): 2.
Posted results (ClinicalTrials.gov registry) · primary outcome public domain (ClinicalTrials.gov)
Time to Death [up to 32 months]. Median, Days (Inter-Quartile Range). 6 mm Zilver (n=64): 132.6 (74 to 328); 10 mm Zilver (n=88): 186 (92 to 304); 10 mm Wallstent (n=89): 170 (99 to 273). Analysis, 6 mm Zilver vs 10 mm Zilver vs 10 mm Wallstent: p=.32; Log Rank; superiority or other. Analysis, 6 mm Zilver vs 10 mm Zilver vs 10 mm Wallstent: p=.69; Kruskal-Wallis; superiority or other.
Papers 1
- Registry-listed results paperComparative performance of uncoated, self-expanding metal biliary stents of different designs in 2 diameters: final res… DOI 10.1016/j.gie.2008.11.018↗
The registry lists these as results papers; in a hand check of 27 such links about 78% reported the trial’s results, so read the paper before treating it as the outcome.