Pancreatic Cancer Research

Clinical trialNo phase (not a drug trial)Completed

Comparison of Duodenoscope With Single-use Distal Cover and the Conventional Reusable Duodenoscope

Comparison of Outcomes in Patients Undergoing Endoscopic Retrograde Cholangiopancreatoscopy (ERCP) Using Duodenoscope With Single-use Distal Cover and Conventional Reusable Duodenoscope: A Randomized Controlled Trial

Outcome
Primary endpoint: no statistically significant difference (p = 0.8, as posted; posted threshold 0.05)
Phase
No phase (not a drug trial)
Status
Completed
Study type
Interventional
Start
25 October 2022
Ended
29 July 2024
Enrollment
319 participants
Sites
1
Primary outcomes
Number of Participants With Procedure-related Adverse Events
Sponsor type
Academic / other
Results posted
15 September 2025
Last update
15 September 2025

These fields come from the trial’s registry entry. A summary of its population, design, endpoints and results, each tied to a quote, has not been written for this trial yet.

Results

Primary endpoint
Primary endpoint: no statistically significant difference (p = 0.8, as posted; posted threshold 0.05)As posted on ClinicalTrials.gov ↗, registry record of 15 September 2025
Primary outcome
Number of Participants With Procedure-related Adverse EventsDuodenoscope With Disposable Distal End Caps: 8 of 159Conventional Reusable Duodenoscope: 9 of 160Duodenoscope With Disposable Distal End Caps vs Conventional Reusable Duodenoscope: p = 0.8; Chi-squared (not statistically significant against the posted threshold 0.05; outcome 1, analysis 1)Values: ParticipantsTime frame: 3 months
Participants
Duodenoscope With Disposable Distal End Caps: 159 started, 159 completedConventional Reusable Duodenoscope: 160 started, 160 completed
Serious adverse events
Duodenoscope With Disposable Distal End Caps: 8 of 159Conventional Reusable Duodenoscope: 9 of 160Participants with at least one serious adverse event, of those at risk, as posted
12 secondary outcomes, the first 8 shown
Duodenoscope Maneuverability: Intubation Into Esophagus
Rating = 1: Duodenoscope With Disposable Distal End… 153/159; Conventional Reusable Duodenoscope 154/160 [Participants] (first of 5 rows) Duodenoscope With Disposable Distal End… vs Conventional Reusable Duodenoscope: p = 1; Fisher Exact
Duodenoscope Maneuverability: Passage Into Stomach
Rating = 1: Duodenoscope With Disposable Distal End… 153/159; Conventional Reusable Duodenoscope 155/160 [Participants] (first of 5 rows) Duodenoscope With Disposable Distal End… vs Conventional Reusable Duodenoscope: p = 0.7; Fisher Exact
Duodenoscope Maneuverability: Navigation Across Pylorus
Rating = 1: Duodenoscope With Disposable Distal End… 151/159; Conventional Reusable Duodenoscope 155/160 [Participants] (first of 5 rows) Duodenoscope With Disposable Distal End… vs Conventional Reusable Duodenoscope: p = 0.4; Fisher Exact
Duodenoscope Maneuverability: Achieving Short Position of the Scope
Rating = 1: Duodenoscope With Disposable Distal End… 154/159; Conventional Reusable Duodenoscope 155/160 [Participants] (first of 5 rows) Duodenoscope With Disposable Distal End… vs Conventional Reusable Duodenoscope: p = 0.8; Fisher Exact
Duodenoscope Maneuverability: Achieving Optimal Papillary Position
Rating = 1: Duodenoscope With Disposable Distal End… 154/159; Conventional Reusable Duodenoscope 154/160 [Participants] (first of 5 rows) Duodenoscope With Disposable Distal End… vs Conventional Reusable Duodenoscope: p = 0.8; Fisher Exact
Duodenoscope Mechanical Characteristics Profile: Scope Stiffness
Rating = 1: Duodenoscope With Disposable Distal End… 155/159; Conventional Reusable Duodenoscope 155/160 [Participants] (first of 5 rows) Duodenoscope With Disposable Distal End… vs Conventional Reusable Duodenoscope: p = 0.7; Fisher Exact
Duodenoscope Mechanical Characteristics Profile: Air-water Button Functionality
Rating = 1: Duodenoscope With Disposable Distal End… 155/159; Conventional Reusable Duodenoscope 156/160 [Participants] (first of 5 rows) Duodenoscope With Disposable Distal End… vs Conventional Reusable Duodenoscope: p = 0.6; Fisher Exact
Duodenoscope Mechanical Characteristics Profile: Elevator Efficiency
Rating = 1: Duodenoscope With Disposable Distal End… 157/159; Conventional Reusable Duodenoscope 158/160 [Participants] (first of 5 rows) Duodenoscope With Disposable Distal End… vs Conventional Reusable Duodenoscope: p = 0.5; Fisher Exact

Every value is copied from the trial’s results record at ClinicalTrials.gov. “Statistically significant” describes the posted test only: it does not say which group did better, and nothing here is read from a paper.

Ask this document

Answers come only from this document’s own text, as highlighted quotes. If it doesn’t say, you are told so instead of given a guess.

Try

Source text

Brief summary public domain (ClinicalTrials.gov)

The purpose of the study is to compare the success rates of procedure success and complication rates in patients undergoing Endoscopic retrograde cholangiopancreatography (ERCP) using two types of currently available endoscopes. These are (1) duodenoscope with a Single-use distal cover or (2) a conventional reusable duodenoscope.

Detailed description public domain (ClinicalTrials.gov)

The risks associated with the ERCP procedure will be explained to patients who meets the eligibility criteria. Consenting patients will be randomized to undergo Endoscopic retrograde cholangiopancreatography (ERCP) using either (1) a duodenoscope with a Single-use distal cover or (2) a conventional reusable duodenoscope. Assessments will be made during, and immediately after the procedure, then patients will be called after 1 week, 1 month, and 3 months after the procedure to get information about adverse events.

Posted results (ClinicalTrials.gov registry) · primary outcome public domain (ClinicalTrials.gov)

Number of Participants With Procedure-related Adverse Events [3 months]. Count Of Participants, Participants. Duodenoscope With Disposable Distal End Caps (n=159): 8; Conventional Reusable Duodenoscope (n=160): 9. Analysis, Duodenoscope With Disposable Distal End Caps vs Conventional Reusable Duodenoscope: p=0.8; Chi-squared; other.

Posted results (ClinicalTrials.gov registry) · secondary outcome public domain (ClinicalTrials.gov)

Duodenoscope Maneuverability: Intubation Into Esophagus [Maneuverability will be assessed only during the procedure (approximately 10 minutes)]. Count Of Participants, Participants. Rating = 1: Duodenoscope With Disposable Distal End Caps (n=159): 153; Conventional Reusable Duodenoscope (n=160): 154. Rating = 2: Duodenoscope With Disposable Distal End Caps (n=159): 5; Conventional Reusable Duodenoscope (n=160): 5. Rating = 3: Duodenoscope With Disposable Distal End Caps (n=159): 1; Conventional Reusable Duodenoscope (n=160): 1. Rating = 4: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Rating = 5: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Analysis, Duodenoscope With Disposable Distal End Caps vs Conventional Reusable Duodenoscope: p=1; Fisher Exact; other.

Posted results (ClinicalTrials.gov registry) · secondary outcome public domain (ClinicalTrials.gov)

Duodenoscope Maneuverability: Passage Into Stomach [Maneuverability will be assessed only during the procedure (approximately 10 minutes)]. Count Of Participants, Participants. Rating = 1: Duodenoscope With Disposable Distal End Caps (n=159): 153; Conventional Reusable Duodenoscope (n=160): 155. Rating = 2: Duodenoscope With Disposable Distal End Caps (n=159): 3; Conventional Reusable Duodenoscope (n=160): 2. Rating = 3: Duodenoscope With Disposable Distal End Caps (n=159): 3; Conventional Reusable Duodenoscope (n=160): 2. Rating = 4: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 1. Rating = 5: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Analysis, Duodenoscope With Disposable Distal End Caps vs Conventional Reusable Duodenoscope: p=0.7; Fisher Exact; other.

Posted results (ClinicalTrials.gov registry) · secondary outcome public domain (ClinicalTrials.gov)

Duodenoscope Maneuverability: Navigation Across Pylorus [Maneuverability will be assessed only during the procedure (approximately 10 minutes)]. Count Of Participants, Participants. Rating = 1: Duodenoscope With Disposable Distal End Caps (n=159): 151; Conventional Reusable Duodenoscope (n=160): 155. Rating = 2: Duodenoscope With Disposable Distal End Caps (n=159): 3; Conventional Reusable Duodenoscope (n=160): 4. Rating = 3: Duodenoscope With Disposable Distal End Caps (n=159): 1; Conventional Reusable Duodenoscope (n=160): 1. Rating = 4: Duodenoscope With Disposable Distal End Caps (n=159): 2; Conventional Reusable Duodenoscope (n=160): 0. Rating = 5: Duodenoscope With Disposable Distal End Caps (n=159): 2; Conventional Reusable Duodenoscope (n=160): 0. Analysis, Duodenoscope With Disposable Distal End Caps vs Conventional Reusable Duodenoscope: p=0.4; Fisher Exact; other.

Posted results (ClinicalTrials.gov registry) · secondary outcome public domain (ClinicalTrials.gov)

Duodenoscope Maneuverability: Achieving Short Position of the Scope [Maneuverability will be assessed only during the procedure (approximately 10 minutes)]. Count Of Participants, Participants. Rating = 1: Duodenoscope With Disposable Distal End Caps (n=159): 154; Conventional Reusable Duodenoscope (n=160): 155. Rating = 2: Duodenoscope With Disposable Distal End Caps (n=159): 3; Conventional Reusable Duodenoscope (n=160): 4. Rating = 3: Duodenoscope With Disposable Distal End Caps (n=159): 2; Conventional Reusable Duodenoscope (n=160): 1. Rating = 4: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Rating = 5: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Analysis, Duodenoscope With Disposable Distal End Caps vs Conventional Reusable Duodenoscope: p=0.8; Fisher Exact; other.

Posted results (ClinicalTrials.gov registry) · secondary outcome public domain (ClinicalTrials.gov)

Duodenoscope Maneuverability: Achieving Optimal Papillary Position [Maneuverability will be assessed only during the procedure (approximately 10 minutes)]. Count Of Participants, Participants. Rating = 1: Duodenoscope With Disposable Distal End Caps (n=159): 154; Conventional Reusable Duodenoscope (n=160): 154. Rating = 2: Duodenoscope With Disposable Distal End Caps (n=159): 4; Conventional Reusable Duodenoscope (n=160): 4. Rating = 3: Duodenoscope With Disposable Distal End Caps (n=159): 1; Conventional Reusable Duodenoscope (n=160): 2. Rating = 4: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Rating = 5: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Analysis, Duodenoscope With Disposable Distal End Caps vs Conventional Reusable Duodenoscope: p=0.8; Fisher Exact; other.

Posted results (ClinicalTrials.gov registry) · secondary outcome public domain (ClinicalTrials.gov)

Duodenoscope Mechanical Characteristics Profile: Scope Stiffness [Mechanical characteristics assessed only during the procedure (approximately 15 minutes)]. Count Of Participants, Participants. Rating = 1: Duodenoscope With Disposable Distal End Caps (n=159): 155; Conventional Reusable Duodenoscope (n=160): 155. Rating = 2: Duodenoscope With Disposable Distal End Caps (n=159): 4; Conventional Reusable Duodenoscope (n=160): 5. Rating = 3: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Rating = 4: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Rating = 5: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Analysis, Duodenoscope With Disposable Distal End Caps vs Conventional Reusable Duodenoscope: p=0.7; Fisher Exact; other.

Posted results (ClinicalTrials.gov registry) · secondary outcome public domain (ClinicalTrials.gov)

Duodenoscope Mechanical Characteristics Profile: Air-water Button Functionality [Mechanical characteristics assessed only during the procedure (approximately 15 minutes)]. Count Of Participants, Participants. Rating = 1: Duodenoscope With Disposable Distal End Caps (n=159): 155; Conventional Reusable Duodenoscope (n=160): 156. Rating = 2: Duodenoscope With Disposable Distal End Caps (n=159): 3; Conventional Reusable Duodenoscope (n=160): 4. Rating = 3: Duodenoscope With Disposable Distal End Caps (n=159): 1; Conventional Reusable Duodenoscope (n=160): 0. Rating = 4: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Rating = 5: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Analysis, Duodenoscope With Disposable Distal End Caps vs Conventional Reusable Duodenoscope: p=0.6; Fisher Exact; other.

Posted results (ClinicalTrials.gov registry) · secondary outcome public domain (ClinicalTrials.gov)

Duodenoscope Mechanical Characteristics Profile: Elevator Efficiency [Mechanical characteristics assessed only during the procedure (approximately 15 minutes)]. Count Of Participants, Participants. Rating = 1: Duodenoscope With Disposable Distal End Caps (n=159): 157; Conventional Reusable Duodenoscope (n=160): 158. Rating = 2: Duodenoscope With Disposable Distal End Caps (n=159): 1; Conventional Reusable Duodenoscope (n=160): 2. Rating = 3: Duodenoscope With Disposable Distal End Caps (n=159): 1; Conventional Reusable Duodenoscope (n=160): 0. Rating = 4: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Rating = 5: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Analysis, Duodenoscope With Disposable Distal End Caps vs Conventional Reusable Duodenoscope: p=0.5; Fisher Exact; other.

Posted results (ClinicalTrials.gov registry) · secondary outcome public domain (ClinicalTrials.gov)

Duodenoscope Mechanical Characteristics Profile: Hand Strain [Mechanical characteristics assessed only during the procedure (approximately 15 minutes)]. Count Of Participants, Participants. Rating = 1: Duodenoscope With Disposable Distal End Caps (n=159): 157; Conventional Reusable Duodenoscope (n=160): 158. Rating = 2: Duodenoscope With Disposable Distal End Caps (n=159): 2; Conventional Reusable Duodenoscope (n=160): 2. Rating = 3: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Rating = 4: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Rating = 5: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Analysis, Duodenoscope With Disposable Distal End Caps vs Conventional Reusable Duodenoscope: p=0.9; Fisher Exact; other.

Posted results (ClinicalTrials.gov registry) · secondary outcome public domain (ClinicalTrials.gov)

Duodenoscope Imaging Characteristics Profile: Image Quality [Imaging characteristics were assessed only during the procedure (approximately 15 minutes)]. Count Of Participants, Participants. Rating = 1: Duodenoscope With Disposable Distal End Caps (n=159): 154; Conventional Reusable Duodenoscope (n=160): 154. Rating = 2: Duodenoscope With Disposable Distal End Caps (n=159): 4; Conventional Reusable Duodenoscope (n=160): 4. Rating = 3: Duodenoscope With Disposable Distal End Caps (n=159): 1; Conventional Reusable Duodenoscope (n=160): 2. Rating = 4: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Rating = 5: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Analysis, Duodenoscope With Disposable Distal End Caps vs Conventional Reusable Duodenoscope: p=0.8; Fisher Exact; other.

Posted results (ClinicalTrials.gov registry) · secondary outcome public domain (ClinicalTrials.gov)

Duodenoscope Imaging Characteristics Profile: Image Stability [Imaging characteristics were assessed only during the procedure (approximately 15 minutes)]. Count Of Participants, Participants. Rating = 1: Duodenoscope With Disposable Distal End Caps (n=159): 156; Conventional Reusable Duodenoscope (n=160): 157. Rating = 2: Duodenoscope With Disposable Distal End Caps (n=159): 3; Conventional Reusable Duodenoscope (n=160): 3. Rating = 3: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Rating = 4: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Rating = 5: Duodenoscope With Disposable Distal End Caps (n=159): 0; Conventional Reusable Duodenoscope (n=160): 0. Analysis, Duodenoscope With Disposable Distal End Caps vs Conventional Reusable Duodenoscope: p=0.9; Fisher Exact; other.

Posted results (ClinicalTrials.gov registry) · secondary outcome public domain (ClinicalTrials.gov)

Number of Participant With Procure Success (Ability to Perform Interventions) [30-60 minutes]. Count Of Participants, Participants. Duodenoscope With Disposable Distal End Caps (n=159): 154; Conventional Reusable Duodenoscope (n=160): 155.